Sign in →

Test Code F8A Coagulation Factor VIII Activity Assay, Plasma

Reporting Name

Coag Factor VIII Activity Assay, P

Useful For

Diagnosing hemophilia A

 

Diagnosing von Willebrand disease when measured with the von Willebrand factor (VWF) antigen and VWF activity

 

Diagnosing acquired deficiency states

 

Investigation of prolonged activated partial thromboplastin time

 

Monitoring infusions of factor VIII replacement during interventional procedures and prophylactic infusions

 

This test is not useful for inferring carrier status in suspected female carriers of hemophilia A, unless it is 50% of normal (<28% activity in adults).

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Specimen Type

Plasma Na Cit


Ordering Guidance


Coagulation testing is highly complex, often requiring the performance of multiple assays and correlation with clinical information. For that reason, consider ordering a Coagulation Consultation.



Necessary Information


If priority specimen, mark request form, give reason, and request a call-back.



Specimen Required


Specimen Type: Platelet-poor plasma

Patient Preparation: Patient must not be receiving Coumadin (warfarin) or heparin therapy.

Collection Container/Tube: Light-blue top (3.2% sodium citrate)

Submission Container/Tube: Plastic vial

Specimen Volume: 1 mL

Collection Instructions:

1. Specimen must be collected prior to factor replacement therapy.

2. For complete instructions, see Coagulation Guidelines for Specimen Handling and Processing.

3. Centrifuge, transfer all plasma into a plastic vial, and centrifuge plasma again.

4. Aliquot plasma into a plastic vial leaving 0.25 mL in the bottom of centrifuged vial.

5. Freeze plasma immediately (no longer than 4 hours after collection) at -20° C or ideally, at or below -40° C.

Additional Information:

1. Double-centrifuged specimen is critical for accurate results as platelet contamination may cause spurious results.

2. Each coagulation assay requested should have its own vial.


Specimen Minimum Volume

0.5 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Plasma Na Cit Frozen 14 days

Day(s) Performed

Monday through Saturday

Test Classification

This test has been modified from the manufacturer's instructions. Its performance characteristics were determined by Mayo Clinic in a manner consistent with CLIA requirements. This test has not been cleared or approved by the US Food and Drug Administration.

CPT Code Information

85240

LOINC Code Information

Test ID Test Order Name Order LOINC Value
F8A Coag Factor VIII Activity Assay, P 3209-4

 

Result ID Test Result Name Result LOINC Value
F8A Coag Factor VIII Activity Assay, P 3209-4

Clinical Information

Factor VIII is synthesized in the liver and, perhaps, in other tissues. It is a coagulation cofactor that circulates bound to von Willebrand factor and is part of the intrinsic coagulation pathway. The biological half-life is 9 to 18 hours (average is 12 hours).

 

Congenital factor VIII deficiency is the cause of hemophilia A, which has an incidence of 1 in 10,000 and is inherited in a recessive sex-linked manner on the X chromosome. Severe deficiency (<1%) characteristically demonstrates as hemarthrosis, deep-tissue bleeding, excessive bleeding with trauma, and ecchymoses.

 

Factor VIII may be decreased in von Willebrand disease. Acquired deficiency states also occur.

 

Antibodies specific for factor VIII are the most commonly occurring specific inhibitors of coagulation factors and can produce serious bleeding disorders (acquired hemophilia).

 

Spuriously decreased results may occur, as factor VIII is highly susceptible to proteolytic inactivation.

Cautions

Factor VIII is a labile protein. Improper handling of a specimen may give a false result.

 

Factor VIII is highly susceptible to proteolytic inactivation, with the potential for spuriously decreased assay results. Normal results can be regarded as reliable, but decreased results need to be correlated with other clinical and laboratory information. Repeat testing may be necessary.

 

Factor VIII activity in frozen-thawed plasma specimens may be 10% to 20% lower than if assayed in fresh specimens, even under optimum conditions of processing and transportation, and may be even lower if these conditions are suboptimal.

 

Factor VIII activity rises in response to a number of factors, including pregnancy, estrogen therapy, stress, disease, etc.

 

Once artefactual reduction of factor VIII is excluded, it is important to measure von Willebrand factor levels to ensure that the patient does not have von Willebrand disease.

Report Available

1 to 3 days

Specimen Retention Time

7 days

Reject Due To

Gross hemolysis Reject
Gross lipemia Reject
Gross icterus Reject

Testing Algorithm

For information see Hemophilia Testing Algorithm.

Method Name

Optical Clot-Based

Forms

If not ordering electronically, complete, print, and send a Coagulation Test Request (T753) with the specimen.